Describe five accidents that have
happened in radiotherapy. Explain the reason for these accidents and what is
your strategy to avoid them?
On 27 February 2001 , a radiobiological accident has happened in
Bialystok Oncology Centre (BOC) in Poland where five patient has been affected
while they undergoing radiotherapy treatment. The patient were given higher
dose than prescribe dose resulting in radiation induced injuries.
Check your essay before you submit. See exactly what your professor sees.
See your AI and plagiarism results before your instructor does.Get the exact same report your professor uses. Trusted by 50,000+ students worldwide.
The cause of the accident there was a
transitory loss of electrical power causing an automatic shutdown of the linear
accelerator. The power cut during the radiotherapy session. After the
recuperation of electrical power , the machine has been check and restarted. The
treatments were resumed and the patient receiving radiotherapy at the time of
the power cut and four more patients were treated.
Two of the patient has experienced itching and
burning during the therapy, leading the staff to stop the treatment,
subsequence of dosimetry measurements shows that there is a significant high
output than expected. Further checked has conducted and shows that the dose
monitoring system of the linac was not working properly, and one of the
interlock was damaged. However, the patients have had different range of
radiation injuries.
In order to prevent such an event to occur in
the future , a quality assurance programmes suitable for the linear accelerator
used in the medical field need to include dosimetry checks of beam output
before the resumption of the patient treatment following the power cut or
unusual event. If the QA was implemented
properly , the interlocks would stop the linac and patient would not overdosed.
The
Instituto Oncológico Nacional (ION) in Panama provides treatment for
cancer patients using radiotherapy. They use blocks of shielding material to
modify the shapes of the radiation beams to protect normal tissue, including
critical structures, during treatment . A computerized treatment planning
system (TPS) was used by the ION to calculate the resulting dose distributions
and determine treatment times. The data for each shielding block should be
entered into the TPS separately. The TPS allows a maximum of four shielding
blocks per field to be taken into account when calculating treatment times and
dose distributions. According to information provided to the IAEA Team, in
order to satisfy the request of a radiation oncologist to include five blocks
in the field, in August 2000 the method of digitizing shielding blocks was
changed. It was found that it was possible to enter data into the TPS for
multiple shielding blocks together as if they were a single block, thereby
apparently overcoming the limitation of four blocks per field. As was found
later, although the TPS accepted entry of the data for multiple shielding
blocks as if they were a single block, at least one of the ways in which the
data were entered the computer output indicated a treatment time substantially
longer than it should have been. The result was that patients received a
proportionately higher dose than that prescribed. The modified treatment
protocol was used for 28 patients, who were treated between August 2000 and
March 2001 for prostate cancer and cancer of the cervix. The modified protocol
was used without a verification test. In spite of the treatment times being
about twice those required for correct treatment, the error went unnoticed.
Some early symptoms of excessive exposure were noted in some of the irradiated
patients. The seriousness, however, was not realized, with the consequence that
the accidental exposure went unnoticed for a number of months. The continued
emergence of these symptoms, however, eventually led to the accidental exposure
being detected. This was in March 2001. As per the request of Panama government
the IAEA sent a team of five medical doctors and two physicists to Panama to
perform a dosimetric and medical assessment of the accidental exposure and a
medical evaluation of the affected patients’ prognosis and treatment . By the
time of the mission eight patients had already died. At least five of the
deaths were probably radiation related. One death was assumed to be cancer
related and in two cases there was not enough information to decide the cause
of death. All 20 surviving patients were examined by the medical team. Most of
the injuries of these patients were related to the bowel, with a number of
patients suffering persistent bloody diarrhoea, necrosis, ulceration and
anaemia. About three quarters of them are expected to develop serious
complications, which in some cases may ultimately prove fatal.
To
shows the importance of QA programme if simple mistake has been made and has
not been noted it may result in a series issues therefore QA programme working
would prevent of mistakes to occur. Furthermore TPS is critical in radiotherapy
and should be include in the quality control and must verify by manual
calculation of the treatment time and dose at selected point. Lastly , any
producer should be written down or if any new producer or changes needed to
tested , approved and documented as part of QA.
In December 1990, a linear accelerator broke down in a hospital in Spain. A technician
from the company was nearby, maintaining a cobalt unit, and was called over to
the accelerator. The technician started repair work the following work day,
recovering the beam, but an instrument on the control panel always indicated
the maximum electron energy (36 MeV), regardless of the selected electron
energy value e.g. 7, 10, 13 MeV. Treatments resumed in a few days. The
technologists observed the discrepancy between the energy selected and the one
indicated on the instrument on the control panel. The interpretation was that
the indicator must have got stuck at 36 MeV, while the selected energy was the
one indicated on the energy selection keyboard. What happened is that a transistor had
short-circuited, and independent of the voltage at the base, a full current was
always fed to the magnet system. This made it possible to get a beam only when
maximum electron energy was used. To get a beam for each of the electron
energies, they had all been adjusted to the maximum energy.
The design of
this accelerator meant that a homogenous field was achieved through scanning of
the electron beam, where the current of the scanning magnet had to match the
selected electron energy. As the electron energy was at its maximum, the
deflection in the scanning magnets was too small and the field thus became
concentrated in the centre. This increased the energy fluence and therefore the
dose. For 7 MeV, the absorbed dose was about 9 times the intended. This resulting
increase in dose was smaller for higher energies and it became nearly unity
when the selected energy coincided with the actual energy. During the 10 days
of faulty operations, 27 patients were treated using electrons with the
equipment. Of the 27 patients, 15 died as a consequence of the overexposure
(most of them within 1 year). Two more died with radiation as a major
contributing factor.
Lessons to be
learned are which include in QA program, Formal
procedures for returning medical equipment after maintenance , formal
procedures for making it mandatory to report returned medical equipment to the
Physics group, before resuming treatment with patients, Consideration of the
need to verify the radiation beam by the Physics group, when the repair might
have affected beam parameters, And Procedure to perform a full review or
investigation when unusual displays or behaviour of the radiotherapy equipment
occurs.
In 1992, a patient was being treated for anal
carcinoma in Indiana, USA, using a high dose rate brachytherapy afterloader,
where the source is attached to a wire that can be extended under remote
control through one or more catheters in succession into the patient. Five
catheters had been placed into the target volume, and a pre-treatment check using
a dummy wire testing the catheters had been completed without problems. When
the source was being introduced into the catheters, this went well for the
first four of these. Upon attempting to direct the source into the fifth
catheter, the control console reported an error. After several attempts, the
treatment was abandoned.
When the
treatment was terminated, the staff entered the treatment room, disconnected
the HDR unit from the implanted catheters and removed the patient. An area
radiation alarm indicated high radiation levels, but this was ignored. The
staff later reported that the alarm “often malfunctioned” and they
were used to ignore it. A radiation survey meter was available but was not used
to confirm or rule out the area alarm’s signal. The HDR console reported that
the source was “safe” and the patient was transported back to her
nursing home. The hospital staff did not recognize that the source had broken
loose from the guide wire, and had remained inside the catheter.
The catheters
remained in the patient, with the HDR source, as the patient was transported
back to the nursing home. The catheter containing the source fell out four days
later and was placed in a “medical biohazard” trash bag. A radiation
detector identified radiation emissions from the trash in a trailer sometime
later, and the source could be traced back to its origin. The patient had
received a massive overdose, and died shortly after the source fell out.
Lessons
learned for health professionals:
- Ensure
that all staff are:
- Properly trained in radiation safety
procedures; - Properly trained in the operation of
equipment; - Properly trained for emergency situations.
- Properly trained in radiation safety
- Include
in the Quality Assurance programme:
- Formal procedures for verifying the proper
operation of the HDR remote afterloading equipment before patient
treatments; - Formal procedures for verifying the operation
of radiation safety equipment; - Formal procedures for using radiation safety
equipment when radioactive materials are used for therapy.
- Formal procedures for verifying the proper
- Perform
surveys of HDR patients to ensure that the source has returned properly to
the shield after treatment;
A
cobalt unit
for teletherapy at a hospital in Ohio was initially calibrated correctly.
During the period 1974-1976 the physicist failed to perform regular
measurements (calibrations and QA), but relied on estimations of the decay of
the source to predict the dose rate and calculate the treatment time. Rather
than calculating the decay, the physicist plotted the dose rate on a graph
paper and extrapolated the decaying dose rate over time. Decay was determined
from a straight-line plot on a semi-log graph paper with a calendar ordinate.
However, the physicist continued the plot on a page that had linear scales on
both axes.
This created
two problems:
- the
linear Y-axis did not correspond to the original log Y-axis, so straight
line extrapolation resulted in ever-more incorrect output values; - the
linear X-axis did not correspond to the original calendar axis, so
extrapolation led to incorrect date values.
These errors
in predicting the dose-rate were made by the physicist in the time period
1974-1976. The errors resulted in the dose-rate being under-estimated by 10% to
45%, which translates to the patients receiving corresponding overdoses of 10%
to 55% (where the magnitude of error increased almost linearly with time).
The incident came to light because patients started exhibiting symptoms of
overexposure. When the accident was investigated, the physicist produced ten
falsified calibration documents showing the correct machine output.. The output
of the cobalt unit had, in fact, not been checked for 22 months. 426 patients
received significant overdoses as a result.
Lessons
learned for health professionals:
- Provide
independent check of physicist’s safety critical work; - Institute
formal procedures for calibrating a treatment unit on a regular
schedule; - Ensure
that records accurately document performance of accepted QA procedures.
REFERNCES :
Thulani Nyathi, Christopher Colyer, Anup Kumar Bhardwaj,
James Rijken, Jason Morton (2016). Post-upgrade testing on a radiotherapy
oncology information system with an embedded record and verify system following
the IAEA Human Health Report No. 7 recommendations. Phys Med. 2016
Jun; 32(6):
854–858. Published online 2016 May
27. doi: 10.1016/j.ejmp.2016.05.061
[3 marks] 15- What are some major causes of accidents in radiotherapy
due to the R&V systems? Who can they be avoided.
Record and verify systems (RVSs) were
initially developed to reduce the risk of treatment errors, where the treatment
parameters used for a given fraction were set manually and could differ from
the ‘prescribed’ (or ‘intended’) parameters . These parameters, which were to
be applied during the whole treatment course, were obtained as a result of the
treatment planning process, either from a real simulation performed with a
simulator, including skin marking and image verification, or from a computer
based simulation performed on a treatment planning system (TPS), including
computation of the provisional dose distribution.
the RVS must have stored a complete set of
information, including the patient’s identification, prescription, treatment
plan and field parameters, and must allow this information to be entered and
called up correctly for each treatment. The RVS must also be provided with
details about the treatment equipment, including the coordinate, scale and
angle conventions used, beam energies, available field sizes and other parameters
and limitations. When these parameters are incorrectly entered in the RVS,
systematic treatment errors will occur.
In fact, although the initial intent of RVSs
was to assess and/or reduce the risk of errors, and although RVSs have proven
to be effective in reaching this aim, several accidents, incidents or near
misses have been reported that can be attributed, more or less directly, to the
use of RVSs. According to the experience at Memorial Sloan-Kettening Cancer
Center, there was a 34% decrease in the frequency of treatment delivery errors
in the transition period between 2001 and 2004, during which time an RVS was
fully implemented, but, in the same time, the errors attributed to the RVS had
increased, representing as much as 55% of all encountered errors.
Some of the most significant types of
reported events related to the use of RVSs are summarized in Table 1. All of
these errors might be partially attributed to a lack of appropriate human
control, since it is perfectly clear that human and organizational factors are
mostly responsible for accidents. In the case of RVSs, it has been further
advocated that the radiation therapists, if not properly informed, could be
naturally inclined to relax their attention due to an ‘excessive reliance’ on
the system, which could contribute to the error rate. Errors are also often due
to a lack of well-defined workflow and procedures. Some other errors might be
due to problems in the system design or implementation. One particular example
is the difficulty of matching regional specifics of language or typography,
such as the presence of accentuated characters or interpretation of the decimal
point, which could result in erroneous data or results. One of the areas of
significant difficulty in the implementation of a new RVS in a clinical
environment is the component that deals with the acceptance testing and
commissioning of such a system. Acceptance testing relates to the evaluation by
the purchaser of a new RVS to ensure that the system meets the specifications
as defined by the user and/or manufacturer. Commissioning relates to the
parameterization and verification that are needed to get the system ready for
clinical use. While acceptance testing is a well-defined and standardized
process for the purchase of other radiation therapy equipment
there currently are no publications that
offer guidance on acceptance testing or commissioning of an RVS.
References
:
IAEA
HUMAN HEALTH REPORTS NO.7 . Record and Verify Systems for Radiation Treatment
of Cancer: Acceptance Testing, Commissioning and Quality Control.
[2 marks] 16- Explain the MV Imager calibration procedure in
detail.
Recently, the role of EPIDs has been expanded
beyond patient imaging to become a useful
tool for radiotherapy
dosimetry. Conventional methods for calibration of EPID
includes dark-field and flood-field irradiations at a fixed panel position and
requires a beam profile correction for off-axis ratio differences, which is
typically collected with a scanning water tank. It is expected that some
variations in detector’s sensitivities are inevitable.
For all image acquisitions,
appropriate imaging software should be used to control incoming signal
data. The software should be capable to
demonstrate the functions of the RID (Radiation Image Detector). It detects
automatically the size of the sensor and receives images of the detector in a
16 bit digitized data format (65535 steps). The images are displayed on the
screen with 8 bit resolution. The detector is connected to the computer, which
provides synchronization between the EPID and the Linac. The detector
triggering, is controlled by the Detector Control Board (DCB) which uses the
gun pulse control board to synchronize the reading of data from the detector,
so that image data is read between radiation pulses. When the detector is
triggered, a single image data packet is sent directly from the detector panel
to the frame grabber in the PC, where appropriate software packages for signal
processing, automatically apply a set of corrections to all images, including
offset, gain and bad pixel corrections.
The Offset correction image
(dark field – DF) is used to take out the specific dark current of the detector
(leakage current) or other fixed signals. The Offset image is created and stored
when the detector is not exposed to the radiation, and is then used to correct
all subsequent images.
Sensitivity variations from
pixel to pixel are taken into account by the Gain correction image (flood field
– FF). The value of each pixel in the offset corrected image is divided by the
corresponding pixel value of the normalized Gain to correct also the image dose
calibration under different energy conditions. The processed and corrected
image is performed
by the application of equation 1, where a raw image is acquired by the EPID and subtracted by the
dark field image divided by the normalized Flood Field image ( . is the mean
value of the , used for
image display.
The bad pixel map correction
allows the ‘software repair’ of defected or non-responding pixels to enhance
image quality. It is used to identify pixels that the response cannot correct
for using gain or offset. The corresponding pixel values are set to the mean
value of the neighbouring pixels.
Reference :
aSi
EPID Panel damage. Chapter 3 Material and methods . Image acquisition and
calibration ( online through
Describe five accidents that have happened in radiotherapy. Explain the reason for these accidents and what is your strategy to avoid them? On 27 February 2001 , a radiobiological accident has happened in Bialystok Oncology Centre (BOC) in Poland where five patient has been affected while they undergoing radiotherapy treatment. The patient were given higher dose than prescribe dose resulting in radiation induced injuries. The cause of the accident there was a transitory loss of electrical power causing an automatic shutdown of the linear accelerator. The power cut during the radiotherapy session. After the recuperation of electrical power , the machine has been check and restarted. The treatments were resumed and the patient receiving radiotherapy at the time of the power cut and four more patients were treated. Two of the patient has experienced itching and burning during the therapy, leading the staff to stop the treatment, subsequence of dosimetry measurements shows that there is a significant high output than expected. Further checked has conducted and shows that the dose monitoring system of the linac was not working properly, and one of the interlock was damaged. However, the patients have had different range of radiation injuries. In order to prevent such an event to occur in the future , a quality assurance programmes suitable for the linear accelerator used in the medical field need to include dosimetry checks of beam output before the resumption of the patient treatment following the power cut or unusual event. If the QA was implemented properly , the interlocks would stop the linac and patient would not overdosed. The Instituto Oncológico Nacional (ION) in Panama provides treatment for cancer patients using radiotherapy. They use blocks of shielding material to modify the shapes of the radiation beams to protect normal tissue, including critical structures, during treatment . A computerized treatment planning system (TPS) was used by the ION to calculate the resulting dose distributions and determine treatment times. The data for each shielding block should be entered into the TPS separately. The TPS allows a maximum of four shielding blocks per field to be taken into account when calculating treatment times and dose distributions. According to information provided to the IAEA Team, in order to satisfy the request of a radiation oncologist to include five blocks in the field, in August 2000 the method of digitizing shielding blocks was changed. It was found that it was possible to enter data into the TPS for multiple shielding blocks together as if they were a single block, thereby apparently overcoming the limitation of four blocks per field. As was found later, although the TPS accepted entry of the data for multiple shielding blocks as if they were a single block, at least one of the ways in which the data were entered the computer output indicated a treatment time substantially longer than it should have been. The result was that patients received a proportionately higher dose than that prescribed. The modified treatment protocol was used for 28 patients, who were treated between August 2000 and March 2001 for prostate cancer and cancer of the cervix. The modified protocol was used without a verification test. In spite of the treatment times being about twice those required for correct treatment, the error went unnoticed. Some early symptoms of excessive exposure were noted in some of the irradiated patients. The seriousness, however, was not realized, with the consequence that the accidental exposure went unnoticed for a number of months. The continued emergence of these symptoms, however, eventually led to the accidental exposure being detected. This was in March 2001. As per the request of Panama government the IAEA sent a team of five medical doctors and two physicists to Panama to perform a dosimetric and medical assessment of the accidental exposure and a medical evaluation of the affected patients’ prognosis and treatment . By the time of the mission eight patients had already died. At least five of the deaths were probably radiation related. One death was assumed to be cancer related and in two cases there was not enough information to decide the cause of death. All 20 surviving patients were examined by the medical team. Most of the injuries of these patients were related to the bowel, with a number of patients suffering persistent bloody diarrhoea, necrosis, ulceration and anaemia. About three quarters of them are expected to develop serious complications, which in some cases may ultimately prove fatal. To shows the importance of QA programme if simple mistake has been made and has not been noted it may result in a series issues therefore QA programme working would prevent of mistakes to occur. Furthermore TPS is critical in radiotherapy and should be include in the quality control and must verify by manual calculation of the treatment time and dose at selected point. Lastly , any producer should be written down or if any new producer or changes needed to tested , approved and documented as part of QA. In December 1990, a linear accelerator broke down in a hospital in Spain. A technician from the company was nearby, maintaining a cobalt unit, and was called over to the accelerator. The technician started repair work the following work day, recovering the beam, but an instrument on the control panel always indicated the maximum electron energy (36 MeV), regardless of the selected electron energy value e.g. 7, 10, 13 MeV. Treatments resumed in a few days. The technologists observed the discrepancy between the energy selected and the one indicated on the instrument on the control panel. The interpretation was that the indicator must have got stuck at 36 MeV, while the selected energy was the one indicated on the energy selection keyboard. What happened is that a transistor had short-circuited, and independent of the voltage at the base, a full current was always fed to the magnet system. This made it possible to get a beam only when maximum electron energy was used. To get a beam for each of the electron energies, they had all been adjusted to the maximum energy. The design of this accelerator meant that a homogenous field was achieved through scanning of the electron beam, where the current of the scanning magnet had to match the selected electron energy. As the electron energy was at its maximum, the deflection in the scanning magnets was too small and the field thus became concentrated in the centre. This increased the energy fluence and therefore the dose. For 7 MeV, the absorbed dose was about 9 times the intended. This resulting increase in dose was smaller for higher energies and it became nearly unity when the selected energy coincided with the actual energy. During the 10 days of faulty operations, 27 patients were treated using electrons with the equipment. Of the 27 patients, 15 died as a consequence of the overexposure (most of them within 1 year). Two more died with radiation as a major contributing factor. Lessons to be learned are which include in QA program, Formal procedures for returning medical equipment after maintenance , formal procedures for making it mandatory to report returned medical equipment to the Physics group, before resuming treatment with patients, Consideration of the need to verify the radiation beam by the Physics group, when the repair might have affected beam parameters, And Procedure to perform a full review or investigation when unusual displays or behaviour of the radiotherapy equipment occurs. In 1992, a patient was being treated for anal carcinoma in Indiana, USA, using a high dose rate brachytherapy afterloader, where the source is attached to a wire that can be extended under remote control through one or more catheters in succession into the patient. Five catheters had been placed into the target volume, and a pre-treatment check using a dummy wire testing the catheters had been completed without problems. When the source was being introduced into the catheters, this went well for the first four of these. Upon attempting to direct the source into the fifth catheter, the control console reported an error. After several attempts, the treatment was abandoned. When the treatment was terminated, the staff entered the treatment room, disconnected the HDR unit from the implanted catheters and removed the patient. An area radiation alarm indicated high radiation levels, but this was ignored. The staff later reported that the alarm “often malfunctioned” and they were used to ignore it. A radiation survey meter was available but was not used to confirm or rule out the area alarm’s signal. The HDR console reported that the source was “safe” and the patient was transported back to her nursing home. The hospital staff did not recognize that the source had broken loose from the guide wire, and had remained inside the catheter. The catheters remained in the patient, with the HDR source, as the patient was transported back to the nursing home. The catheter containing the source fell out four days later and was placed in a “medical biohazard” trash bag. A radiation detector identified radiation emissions from the trash in a trailer sometime later, and the source could be traced back to its origin. The patient had received a massive overdose, and died shortly after the source fell out. Lessons learned for health professionals: Ensure that all staff are: Properly trained in radiation safety procedures; Properly trained in the operation of equipment; Properly trained for emergency situations. Include in the Quality Assurance programme: Formal procedures for verifying the proper operation of the HDR remote afterloading equipment before patient treatments; Formal procedures for verifying the operation of radiation safety equipment; Formal procedures for using radiation safety equipment when radioactive materials are used for therapy. Perform surveys of HDR patients to ensure that the source has returned properly to the shield after treatment; A cobalt unit for teletherapy at a hospital in Ohio was initially calibrated correctly. During the period 1974-1976 the physicist failed to perform regular measurements (calibrations and QA), but relied on estimations of the decay of the source to predict the dose rate and calculate the treatment time. Rather than calculating the decay, the physicist plotted the dose rate on a graph paper and extrapolated the decaying dose rate over time. Decay was determined from a straight-line plot on a semi-log graph paper with a calendar ordinate. However, the physicist continued the plot on a page that had linear scales on both axes. This created two problems: the linear Y-axis did not correspond to the original log Y-axis, so straight line extrapolation resulted in ever-more incorrect output values; the linear X-axis did not correspond to the original calendar axis, so extrapolation led to incorrect date values. These errors in predicting the dose-rate were made by the physicist in the time period 1974-1976. The errors resulted in the dose-rate being under-estimated by 10% to 45%, which translates to the patients receiving corresponding overdoses of 10% to 55% (where the magnitude of error increased almost linearly with time). The incident came to light because patients started exhibiting symptoms of overexposure. When the accident was investigated, the physicist produced ten falsified calibration documents showing the correct machine output.. The output of the cobalt unit had, in fact, not been checked for 22 months. 426 patients received significant overdoses as a result. Lessons learned for health professionals: Provide independent check of physicist’s safety critical work; Institute formal procedures for calibrating a treatment unit on a regular schedule; Ensure that records accurately document performance of accepted QA procedures. REFERNCES : Thulani Nyathi, Christopher Colyer, Anup Kumar Bhardwaj, James Rijken, Jason Morton (2016). Post-upgrade testing on a radiotherapy oncology information system with an embedded record and verify system following the IAEA Human Health Report No. 7 recommendations. Phys Med. 2016 Jun; 32(6): 854–858. Published online 2016 May 27. doi: 10.1016/j.ejmp.2016.05.061 [3 marks] 15- What are some major causes of accidents in radiotherapy due to the R&V systems? Who can they be avoided. Record and verify systems (RVSs) were initially developed to reduce the risk of treatment errors, where the treatment parameters used for a given fraction were set manually and could differ from the ‘prescribed’ (or ‘intended’) parameters . These parameters, which were to be applied during the whole treatment course, were obtained as a result of the treatment planning process, either from a real simulation performed with a simulator, including skin marking and image verification, or from a computer based simulation performed on a treatment planning system (TPS), including computation of the provisional dose distribution. the RVS must have stored a complete set of information, including the patient’s identification, prescription, treatment plan and field parameters, and must allow this information to be entered and called up correctly for each treatment. The RVS must also be provided with details about the treatment equipment, including the coordinate, scale and angle conventions used, beam energies, available field sizes and other parameters and limitations. When these parameters are incorrectly entered in the RVS, systematic treatment errors will occur. In fact, although the initial intent of RVSs was to assess and/or reduce the risk of errors, and although RVSs have proven to be effective in reaching this aim, several accidents, incidents or near misses have been reported that can be attributed, more or less directly, to the use of RVSs. According to the experience at Memorial Sloan-Kettening Cancer Center, there was a 34% decrease in the frequency of treatment delivery errors in the transition period between 2001 and 2004, during which time an RVS was fully implemented, but, in the same time, the errors attributed to the RVS had increased, representing as much as 55% of all encountered errors. Some of the most significant types of reported events related to the use of RVSs are summarized in Table 1. All of these errors might be partially attributed to a lack of appropriate human control, since it is perfectly clear that human and organizational factors are mostly responsible for accidents. In the case of RVSs, it has been further advocated that the radiation therapists, if not properly informed, could be naturally inclined to relax their attention due to an ‘excessive reliance’ on the system, which could contribute to the error rate. Errors are also often due to a lack of well-defined workflow and procedures. Some other errors might be due to problems in the system design or implementation. One particular example is the difficulty of matching regional specifics of language or typography, such as the presence of accentuated characters or interpretation of the decimal point, which could result in erroneous data or results. One of the areas of significant difficulty in the implementation of a new RVS in a clinical environment is the component that deals with the acceptance testing and commissioning of such a system. Acceptance testing relates to the evaluation by the purchaser of a new RVS to ensure that the system meets the specifications as defined by the user and/or manufacturer. Commissioning relates to the parameterization and verification that are needed to get the system ready for clinical use. While acceptance testing is a well-defined and standardized process for the purchase of other radiation therapy equipment there currently are no publications that offer guidance on acceptance testing or commissioning of an RVS. References : IAEA HUMAN HEALTH REPORTS NO.7 . Record and Verify Systems for Radiation Treatment of Cancer: Acceptance Testing, Commissioning and Quality Control. [2 marks] 16- Explain the MV Imager calibration procedure in detail. Recently, the role of EPIDs has been expanded beyond patient imaging to become a useful tool for radiotherapy dosimetry. Conventional methods for calibration of EPID includes dark-field and flood-field irradiations at a fixed panel position and requires a beam profile correction for off-axis ratio differences, which is typically collected with a scanning water tank. It is expected that some variations in detector’s sensitivities are inevitable. For all image acquisitions, appropriate imaging software should be used to control incoming signal data. The software should be capable to demonstrate the functions of the RID (Radiation Image Detector). It detects automatically the size of the sensor and receives images of the detector in a 16 bit digitized data format (65535 steps). The images are displayed on the screen with 8 bit resolution. The detector is connected to the computer, which provides synchronization between the EPID and the Linac. The detector triggering, is controlled by the Detector Control Board (DCB) which uses the gun pulse control board to synchronize the reading of data from the detector, so that image data is read between radiation pulses. When the detector is triggered, a single image data packet is sent directly from the detector panel to the frame grabber in the PC, where appropriate software packages for signal processing, automatically apply a set of corrections to all images, including offset, gain and bad pixel corrections. The Offset correction image (dark field – DF) is used to take out the specific dark current of the detector (leakage current) or other fixed signals. The Offset image is created and stored when the detector is not exposed to the radiation, and is then used to correct all subsequent images. Sensitivity variations from pixel to pixel are taken into account by the Gain correction image (flood field – FF). The value of each pixel in the offset corrected image is divided by the corresponding pixel value of the normalized Gain to correct also the image dose calibration under different energy conditions. The processed and corrected image I(I,j) is performed by the application of equation 1, where a raw image (I_row (i,j)) is acquired by the EPID and subtracted by the dark field image (DF (i,j)) divided by the normalized Flood Field image ( (FF (i,j)). 〖FF〗_mean is the mean value of the (FF (i,j)), used for image display. I(I,j)=((I_row (i,j))-(DF (i,j)))/((FF (i,j)) ) The bad pixel map correction allows the ‘software repair’ of defected or non-responding pixels to enhance image quality. It is used to identify pixels that the response cannot correct for using gain or offset. The corresponding pixel values are set to the mean value of the neighbouring pixels. Reference : aSi EPID Panel damage. Chapter 3 Material and methods . Image acquisition and calibration ( online through https://repositorio.ul.pt/bitstream/10451/1277/7/19191_ulfc091304_tm_7_Chapter_3.pdf) essay Essay Fix.
Struggling with paper writing? Look no further, as you have found the ideal paper writing company! We are a reputable essay writing service that offers high-quality papers at affordable prices. On our user-friendly website, you can request a wide range of assignments. Rest assured that our work is entirely original. Each essay is crafted from scratch, tailored to meet the precise requirements of your assignment. We guarantee that it will successfully pass any plagiarism check.
Get Your Assignments Completed by Expert Writers. Hire Essay Helpers for Any Task
Order essays, term papers, research papers, reaction paper, research proposal, capstone project, discussion, projects, case study, speech/presentation, article, article critique, coursework, book report/review, movie review, annotated bibliography, or another assignment without having to worry about its originality – we offer 100% original content written completely from scratch